Controls
IPAP |
50 cm H₂O
|
CPAP |
4 – 20 cmH20 (passive leak port circuits) cm H₂O
|
EPAP/PEEP (a) |
0 – 25 cmH20 (active valve circuits) cm H₂O
|
EPAP/PEEP (b) |
4 – 25 cmH20 (passive leak port circuits) cm H₂O
|
Pressure support |
0 – 30 cmH2O cm H₂O
|
Tidal volume |
50 – 2000 ml
|
Breath rate(a) |
0 – 60 BPM (AC mode) beats per minute
|
Breath rate(b) |
1 – 60 BPM (all other modes) beats per minute
|
Inspiratory time |
0.3 – 5.0 s Synchrony features
|
Rise time |
1 – 6 (relative scale)
|
Ramp start pressure (a) |
0 – 25 cmH20 (active circuits) cm H₂O
|
Ramp start pressure (b) |
4 – 25 cmH20 (passive circuits) cm H₂O
|
Ramp start pressure (c) |
4 – 19 cmH20 (CPAP mode) cm H₂O
|
Ramp time Off |
5 – 45 min
|
C-Flex Off |
1 – 3 (relative scale)
|
Flow trigger sensitivity |
1 – 9 l/min
|
Flow cycle sensitivity |
10 – 90 %
|
Environmental
Operating temperature |
5 – 40 °C
|
Storage temperature |
-20 – 60 °C
|
Relative humidity |
15 – 95 %
|
Atmospheric pressure |
60 – 110 kPa (450 – 825 mmHg)
|
Patient types
Adult |
Yes
|
Pediatric |
≥ 5 kg or greater
|
Electrical
Input voltage |
100 – 240 VAC and 50/60 Hz and 2.1 A
|
Detachable battery voltage |
14.4 VDC
|
Internal battery life |
3 hours under normal conditions hr
|
Detachable battery life |
3 hours under normal conditions hr
|
external battery connection |
12 VDC
|
Volume modes
Pressure control (PC) |
Yes
|
MPV Mouthpiece Ventilation Mode |
Yes
|
Synchronized intermittent mandatory ventilation (SIMV) |
Yes
|
SIMV with pressure support (SIMV w/PS) |
Yes
|
Control ventilation (CV) |
Yes
|
Alarms
Circuit disconnect Off |
10 – 60 sec
|
Low tidal volume Off |
50 – 2000 ml
|
Apnea Off |
10 – 60 s and 4 – 60 BPM
|
High tidal volume Off |
50 – 2000 ml
|
High minute ventilation Off |
1 – 99 l/min Off, 1 – 99 l/min
|
Low minute ventilation Off |
1 – 99 l/min
|
High respiratory rate Off |
4 – 80 beats per minute
|
Low respiratory rate Off |
4 – 80 beats per minute
|
Pressure modes
Pressure control (PC) |
Yes
|
Pressure control-SIMV (PC-SIMV) |
Yes
|
Spontaneous ventilation (S) |
Yes
|
Spontaneous ventilation with timed back-up (S/T) |
Yes
|
Timed ventilation (T) |
Yes
|
Continuous positive airway pressure (CPAP) |
Yes
|
Average volume assured pressure support (AVAPS) |
With passive circuit type, S, S/T, PC, and T modes only
|
Options
Detachable back-up battery |
Offering up to 3 hours additional operating time
|
Hospital roll stand |
Provides convenient accessory basket and humidifier mount
|
DirectView |
1 GB SD card data storage integrated into the ventilator software
|
Monitored parameters
Tidal volume |
0 – 2000 ml
|
Minute ventilation |
0 – 99 l/min
|
Estimated leak rate |
0 – 200 l/min
|
Respiratory rate |
0 – 80 BPM 0 – 80 beats per minute
|
Peak inspiratory flow |
0 – 200 l/min
|
Peak inspiratory pressure |
0 – 99 cm H₂O
|
Mean airway pressure |
0 – 99 cm H₂O
|
% patient triggered breaths |
0 – 100 %
|
I:E ratio |
9.9:1 – 1:9.9
|
Synchrony features
Auto-Trak sensitivity |
Auto-Trak sensitivity Auto-adaptive triggering, cycling, and leak compensation (available in all modes, passive circuit only)
|
Adjustable flow triggering |
1 – 9 l/min (available in all modes and circuit types) l/min
|
Circuit types
Active exhalation valve with proximal airway pressure (PAP) |
Yes
|
Active exhalation valve with flow sensor |
Yes
|
Passive exhalation port |
Yes
|
Compliance
IEC 60601-1 Medical electrical equipment Part 1: General requirements for safety |
Yes
|
IEC 60601-1-2 General requirements for safety – collateral standard Electromagnetic compatibility – requirements and tests |
Yes
|
IEC 60601-2-12 Medical electrical equipment Part 2-12: Particular requirements for the safety of lung ventilators – Critical Care ventilators |
Yes
|
Oxygen
FiO2 |
21 – 100 %
|
02 flush |
2 min at 100%
|
02 input pressure range |
276 – 600 kPa (40 – 87 psi)
|
madems –